Silicon (Si) X-ray absorption edges
The Si K edge is at 1,839 eV. The tables list 7 absorption edges for silicon; K and L1 lie between 100 eV and 100 keV, the range the lab can scan.
- Atomic number
- Z=14
- Molar mass
- 28.085 g/mol
- Edges in the lab range
- 2 of 7
Open silicon in the lab
Both load silicon as a pure sample with the formula Si. Density is retained from your current settings because these X-ray tables have no density. Enter your sample's density on the Sample page before planning a thickness. To scan another edge, use the links in the table below.
Absorption edges of silicon
| Edge | Energy (eV) | Jump ratio | Fluorescence yield | In the lab |
|---|---|---|---|---|
| K | 1,839 | 10.37 | 0.042911 | ScanSoft X-ray |
| L1 | 149.7 | 1.09471 | 0.00003 | ScanSoft X-ray |
| L2 | 99.8 | 1 | 0.00037 | Outside range |
| L3 | 99.2 | 1 | 0.00038 | Outside range |
| M1 | 8 | 1 | 0.00000588 | Outside range |
| M2 | 2 | 1 | 0 | Outside range |
| M3 | 2 | 1 | 0 | Outside range |
Jump ratio is the photoabsorption just above the edge divided by the photoabsorption just below it. Fluorescence yield is the probability that a core hole in that shell decays by emitting an X-ray. Both are dimensionless. The lab can scan across edges strictly between 100 eV and 100 keV; other edges are listed for reference. A listed edge with jump ratio 1 has no step for thickness planning.
Transmission, fluorescence or electron yield
K and L1 are soft X-ray edges in the lab's classification (below 2 keV). Transmission generally needs thin films or membranes, so the planner suggests fluorescence or electron yield. The fluorescence yield at K and L1 is below 0.05, where the lab suggests considering total electron yield.
These are planning rules of thumb. Sample thickness, self-absorption and the beamline's energy range and detectors still matter. The lab simulates transmission; it does not simulate fluorescence or electron-yield detector counts.
Sample presets containing silicon
- Quartz SiO2, 2.65 g/cm3, opens at the Si K edge
These links load the preset's density and select silicon as the absorber. Check the density source on the Sample page and adjust it for your sample.
Values from XrayDB 9.2 (Elam, Ravel and Sieber, 2002), released into the public domain. These isolated-atom tables do not include chemical shifts, XANES or EXAFS fine structure. Methods and data sources